Dimensions: Diameter : 27 / 21 mm ; Length : 207 mm
Weight: 350 g (sensor + 3 m cable)
Witted materia
Body (electronic part) and clamp in PVC.
Cartridge in DELRIN, Special pH glass, platinum, Inbox
316L (protective sleeve of the temperature probe)
Cable : polyurethane jacket
Steam gland : Poly amide
Patch with active material (black) – DO DISK : Optical
isolation silicon
Safeway
The glass electrode is vulnerable to:
– chemicals (organic solvents, acids and strong
bases, peroxide, hydrocarbons),
– mechanical treatments (impacts).
The redox potential electrode is sensitive to sulphide
adsorption on platinum.
Maximum pressure: 5 bars
IP classification: IP68
Connection: 9 armoured connectors, polyurethane jacket, bare-wires or
waterproof Fisher connector
Sensor cable :Standard : 3, 7 and 15 m (other length on request)
100 m Max. Up to 100 m with junction box.